G210 InviCell

Long Term Incubator

Encapsulating 20 years of incubator development, the G210 InviCell incubator combines optimal embryo growth conditions with advanced user control. It is the ideal incubation solution for a busy clinic.

  • Defining, Creating, Monitoring and constantly sustaining the optimal environment for human embryos
  • Thoughtful design and advanced user control options make everyday work with the G210 effective, safe and comfortable
  • With its compact size, low daily running costs and the option of a stacking system the G210 is the ideal solution for a busy Clinic

Availability

This product is available for sale in selected countries worldwide.

Combining optimal embryo growth conditions with advanced user control

G210-with-features

Key Features of G210 InviCell and G210 InviCell Plus

Space Saving

Ten individual incubation chambers to ensure minimal stress and risk of cross contamination for the embryos. Significant space saving

Heating

Dedicated heating plates with grooves ensure direct heat transfer to the dish

Responsive Gas Control

Set the required specifications to mix CO2 and O2. Uses pure CO2 and N2 (optional)

Optimal Environment

Embryo conditions, including pH levels, are kept completely stable through advanced and accurate software control of gas and temperature thus providing stress free incubation

Ambient humidity

The G210 is an incubator with ambient humidity which strongly minimizes the risk of fungal growth inside the incubator and makes it easier to clean

Safe

Electro-magnetic fields eliminated by non-inductive EM Neutra™ technology (patented)

Superior Filtering and Hygiene

High efficiency particle filter to remove airborne particles. VOC filter to remove organic compounds

Tracking and Alarm

Data logging and alerts with K-Link™ extensive ethernet based monitoring

Advanced User Control

Advanced intuitive user interface with dedicated security system

Economical

Low running costs because of low gas consumption and use of pure gas

Independent External monitoring

An option for independent external monitoring of key parameters including temperature and gas concentration

Certified

CE Marked

Optimal Environment for Human Embryos

g210

Decreasing embryo stress is a key factor in long term incubation.

The G210 has several characteristics to ensure minimal stress.
Highly stable environment of both gas and temperature
VOC filter, HEPA14 filter and UV light* (*Not available in the USA) ensuring clean air in the chambers
Individual chambers for each patient- 10 in total

Optimal temperature conditions at all times. The advanced heating system of the G210 ensures constant and all-embracing heating in the chambers

  • Each isolated chamber has its own temperature sensor to ensure quick and accurate temperature control of all 6 sides in the chamber
  • The unique non-inductive EM NeutraTM heating system gives no electro-magnetic fields around the embryos
  • Emulate the natural cycle of basal body temperature with the advanced software based heating technology

Embryo conditions including pH levels are kept stabile through the combination of accurate gas and temperature control.

  • Increased accuracy and stability with dedicated gas control algorithms
  • The built-in advanced gas mixer allows gas concentration to be set at the required level
  • Ambient humidity based incubation significantly reduces the risk for fungal growth and cross contamination
  • The unidirectional gas flow system ensures even gas distribution throughout the chambers
  • To optimize recovery time and keep gas consumption low, gas flow is paused when the lid is open

Advanced User Control

Large 7“ touch-screen display gives you intuitive control of all
parameters and on screen data display.

A quick overview of each chamber and key parameters on screen

Enter patient data directly on the screen for each individual. No more writing on the device with markers or pens

Multi-level password protection ensures no unintended access to settings

Graphic display of key parameters over time

React quickly to changes in the incubator with the extensive audible and visible monitoring and alarms. Alarms can easily be transferred to the hospital central monitoring system.

Temperature — Gas — Service needs

Keep Track of Incubator from anywhere in the Clinic

K-Link Ethernet based monitoring system
Monitor all your incubators directly from your office. The K-Link Ethernet based software provides external monitoring of all key parameters in your G210 InviCell. Multiple G210s can be monitored from the same PC.

Parameters continuously monitored with K-Link:
— Temperature per chamber
— CO2 and O2 levels
— Gas pressure and consumption
— Alarms and warnings
— Service requirements

K-Link E-mail alerts
In case of an alarm K-Link can be set up to automatically send an e-mail to a designated address.

Excel Export
All data is automatically exported to spreadsheet for in depth analysis.

The G210 Invicell Plus — Extra Safety Added

G210 InviCell Plus is designed to meet the increasing demands for external independent monitoring

Continuous pH Measurements

pH is a crucial parameter in embryo culturing. With the G210 InviCell Plus, pH can be measured continuously. By reading the pH in one dedicated chamber, you have the pH for all chambers.

The G210 InviCell Plus is based on a built-in connection to the pH Online System by MTG with the following characteristics:

By connecting the G210 InviCell Plus with the pH online System and placing a special pre-calibrated dish with media in the chamber pH is continuously monitored

Infrared monitoring Resolution of 0.01 pH and an accuracy of +- 0.03 pH Range 5.5 – 9.0 pH

incubator-diagram

Independent Temperature Monitoring

The G210 InviCell plus is prepared for individual temperature monitoring of each culture chamber in G210.

Monitor Temperature Sensors are isolated from the Control sensors or any part of the Incubator control system

The system is based on 10 individual Temperature Sensors. One for each chamber.

Sensor interface board is electrically isolated from the G210 Incubator Control system and has its own Power supply

Independent External Gas Monitoring

To perform true independent monitoring, it is critical to measure the exact point prior to the gas entering the chambers.

In the G210 InviCell Plus a special gas system access chamber has been installed to allow for this.

An external sensor can be placed in the gas line just before the chambers.

The sensor can be read independently of the G210.

The sensor is powered independently.

GasMixerChamber

Space saving with the Stacking System for the G210

The unique Stacking System has capacity for 3 incubators

G210 InviCell

Dimensions (WxDxH) mm
860x550x170
Weight
45 kg
Temperature range ambient to
42.9oC
CO₂ / O₂ range
2-10% / 2-20%
Connection to 100% CO2 and 100% N2
Yes
Gas consumption (Typically)
1 l/h CO₂& 5 l/h N₂
Datalogger system — K-Link
Yes
External Alarm Port
Yes
Power consumption
Max: 270 W
Avg: 65 W chamber 37oC
Ambient humidity and temperature
Max 75% RH & 20-30oC
Heating plates
NUNC/Falcon/Vitrolife
Prepared for pH
No
Prepared for external monitoring
No
No of temperature sensors
11

G210 InviCell Plus

Dimensions (WxDxH) mm
860x550x170
Weight
46 kg
Temperature range ambient to
42.9oC
CO₂ / O₂ range
2-10% / 2-20%
Connection to 100% CO2 and 100% N2
Yes
Gas consumption (Typically)
1 l/h CO₂& 5 l/h N₂
Datalogger system — K-Link
Yes
External Alarm Port
Yes
Power consumption
Max: 270 W
Avg: 65 W chamber 37oC
Ambient humidity and temperature
Max 75% RH & 20-30oC
Heating plates
NUNC/Falcon/Vitrolife
Prepared for pH
Yes
Prepared for external monitoring
CO2 + Temperature
No of temperature sensors
21
[download id=»15843″ ]
Part Number Part Description
59500 G210 InviCell
60000 G210 InviCell Plus